Feasibility of a New Diagnostic Device to Assess Small Intestinal Dysbiosis in Routine Clinical Setting.
The purpose of this study is to determine if the gas-sensing capsule can be used to identify increased numbers and a change of the composition of microbiota or microorganisms in the gut to assess small intestinal dysbiosis.
Study details:
In order to diagnose an increased bacterial density in the small intestine (e. g small intestinal bacterial overgrowth) metabolic activity of the microbiota colonising the small intestine is measured by exposing these bacteria with readily fermentable carbohydrates and measure subsequently (e. g.
for 2 hrs) the change of the concentration of hydrogen (H2) and/or methane in the exhaled breath that results from carbohydrate fermentation in the small intestine. This is the principle of existing breath hydrogen tests, but they are poor at distinguishing where in the intestine the H2 is being produced, and their sensitivity is diminished by 'noise' in the breath H2 as a consequence of hydrogen or methane producing bacteria in the colon and this is frequently observed in the routine clinical setting. These shortcomings can be theoretically overcome by using the Atmo Gas CapsuleR which measures after ingestion gas concentrations during gastrointestinal transit in the lumen of the gut.
As a consequence at the site of fermentation parts per hundred can be measured, rather than the parts per million (ppm) when methane or hydrogen are measured in the exhaled breath. An additional advantage is that the location of the capsule is known, and therefore the source of gas production from increased bacterial load can be determined. Preliminary studies using the gas capsule have shown good correlation of patterns of breath H2 with those of regional H2 patterns generated by the gas-sensing capsule.
In fact, the capsule demonstrated far greater sensitivity in measuring H2 production and a vastly superior signal-to-noise ratio in response to a fermentable load than breath testing. Hence, investigating the utility of the gas-sensing capsule as a means for 'direct' assessment of microbial density presents an opportunity to overcome the shortcomings associated with the current breath test.
Eligibility criteria
Researchers look for people who fit a certain description, called eligibility criteria. See if you qualify.
Inclusion criteria
Exclusion criteria
Eligibility
Age eligible for study : 18 and older
Healthy volunteers accepted : No
Gender eligible for study: All
Things to know
Study dates
Study start: 2021-07-14
Primary completion: 2024-06-01
Study completion finish: 2024-08-01
Study type
DIAGNOSTIC
Phase
NA
Trial ID
NCT04910815
Intervention or treatment
DEVICE: Atmo Gas Capsule
DIAGNOSTIC_TEST: Glucose Breath Test
DIAGNOSTIC_TEST: Fructoolifosaccharides (FOS) Breath Test
PROCEDURE: Jejunal Aspiration and culture
DRUG: Rifaximin
DRUG: Placebo
Conditions
- • Small Intestinal Bacterial Overgrowth
Find a site
Closest Location:
Princess Alexandra Hospital
Research sites nearby
Select from list below to view details:
Princess Alexandra Hospital
Woolloongabba, Queensland, Australia
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Participant Group/Arm | Intervention/Treatment |
---|---|
EXPERIMENTAL: Primary recruitment
| DEVICE: Atmo Gas Capsule
|
ACTIVE_COMPARATOR: Active SIBO Arm - Rifaximin
| DEVICE: Atmo Gas Capsule
|
PLACEBO_COMPARATOR: Active SIBO Arm - Placebo
| DEVICE: Atmo Gas Capsule
|
What is the study measuring?
Primary outcome
Primary Outcome Measure | Primary Outcome Description | Primary Outcome Time Frame |
---|---|---|
Structured Assessment of Gastro-Intestinal Symptoms | 22 item assessing gastrointestinal symptoms. Response items are on a 5-point scale from (1 )no problem to (5) very severe problem. This is given to patients attending the Department of Gastroenterology \& Hepatology at Princess Alexandra Hospital as part of their routine clinical care. | 30 minutes prior to endoscopy |
Hydrogen fermentation in the small intestine as measured by the Atmo capsule | Approximately 30 minutes before the endoscopic procedure, participants will be asked to swallow the capsule (Atmo capsule), which is the size of a large vitamin pill. If the capsule still is in the stomach at the time of the procedure it will be endoscopically pushed into the small intestine. 5g of Fructooligosaccharide (FOS) in 25mL of water will be flushed via the working channel of the endoscope into the small intestine to ensure that carbohydrates are available for fermentation. Subsequently the capsule is propelled by the normal peristaltic waves through the whole gut and is excreted after several days with the faeces. During transit the capsule transmits data about the concentration of specific gases (e.g. hydrogen) in the gut to a small receiver outside of the body. | 30 minutes prior to endoscopy |
Small Intestine microbial load as assessed by Jejunal Aspiration and Culture | The density of the bacterial colonisation of mucosal biopsies obtained during endoscpoy will be used to determine the role of the bacterial density | During endoscpoic procedure |
Secondary outcome
Secondary Outcome Measure | Secondary Outcome Description | Secondary Outcome Time Frame |
---|---|---|
FOS Breath samples on the day of the endoscopy | Immediately prior to the endoscopy a baseline breath sample (approximately 400 cc) will be collected. FOS will be administered into the duodenum during endoscopy. After completion of the endoscopic procedure additional samples will be taken every 20 minutes up to 2 hours after the procedure. | Up to 2 hours after the endescopic procedure |
Glucose Breath Test | The patients will be asked to perform (not earlier than at least 8 hours after the endoscopic procedure) at home a standard glucose test consistent with the established clinical protocol for the diagnosis of SIBO (see SOP for Hydrogen Breath Test for Small Intestinal Bacterial Overgrowth). For this test patients will drink 75 g of glucose and breath samples will be taken at baseline and every 20 minutes for 2 hrs. Samples will be analysed for increased hydrogen, methane and carbon dioxide using a Breathtracker digital microlyser (Quintron Inc USA). If bacteria are present in the upper part of the small intestine, a rise in 12ppm of hydrogen over baseline is indicative of bacterial overgrowth. | Not earlier than 12 hours after the endoscopic procedure |